-Scope of Application: neonate, pediatric, adult
-8” color TFT screen, optional touch screen
-Gas supply: O2, N2O, AIR
-Tidal volume: 5~1500 ml
-Up to 6 flow tubes for 3 gases
-Selectatec bar: Dual positions
-Integrated breathing circuit with built-in heater
-Highly visible ACGO
-Ventilator Software: V-CMV, P-CMV, PSV, V-SIMV, P-SIMV, PRVC, Standby, Manual/Spont -Spirometry loop: P-V, P-F, F-V, reference Loop
-Auxiliary power outlets: 3
-Reading lamp: LED lighting Included
-Li-ion Battery: Standard 1 battery, 4800mAh; Optional 2 batteries, 9600mAh
-Standard Configurations: Main unit, one Chinese brand vaporizer, O2 cell, Built-in Li-on battery, Disposable adult anesthesia breathing circuit, O2/N2O/Air gas pipelines, Hospital O2/N2O/Air joints
-Options: Extra vaporizer (Chinese/Drager/Penlon/Northern), Touch screen, Patient monitor, ETCO2, Suction module, Pediatric/Neonate anesthesia breathing circuit, Spare cylinder yoke (O2, N2O), AGSS, Gas monitor module (CO2, AG), etc.
-Scope of Application: Adult, child
-Gas Source: O2: 0.27--0.55Mpa; N2O: 0.27--0.55MPa
-Flow Meter: O2: 0.1--1.0L/min, 1.1--10L/min; N20: 0.1--1.0L/min, 1.1--10L/min; Air: 0.1--10L/min
-O2, N2O Linkage and N2O Stopper: When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure <0.2Mpa, the flow of nitrous oxide would be cut off
-Flow Rate of Rapid Oxygen Supply: 25--75L/min
-Vaporizer: Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer
-Respiratory Circuit: Working mode: all-close, semi-close, semi-open
-APL: 0.5-7kPa
-Respiratory Bellows: Bellows for adults, tidal volume range:0--1500ml; Bellows for chil- dren, tidal volume range:0-300ml
-Display Mode: 10.4” Touch Screen, High-definition TFT color LCD
-Ventilation Mode: IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, IMV, VCV, PCV
-Ventilation Tidal volume: 20--1500ml
-12” Patient Monitor (ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR)
-Option: EtCO2, IBP
-Scope of Application: Adult, child
-Gas Source: O2: 0.27--0.55Mpa; N2O: 0.27--0.55MPa
-Flow Meter: O2: 0.1--1.0L/min, 1.1--10L/min; N20: 0.1--1.0L/min, 1.1--10L/min
-O2, N2O Linkage and N2O Stopper: When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure < 0.2 MPa, the flow of nitrous oxide would be cut off
-Flow Rate of Rapid Oxygen Supply: 25--75L/min
-Vaporizer: Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer
-Respiratory Circuit: Working mode: all-close, semi-close, semi-open
-Releasing pressure: 0.5-7kPa
-Respiratory Bellows: Bellows for adults, bellows for children, tidal volume range:0-1500 ml -Display Mode: 10.4” Touch Screen, High-definition TFT color LCD
-Ventilation Mode: IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH (IMV, VCV, PCV)
-Ventilation Tidal volume: 50--1500ml
-Scope of Application: Adult, child
-Gas Source: O2 0.27--0.55MPa
-Flow Meter: O2: 0.1--1.0L/min, 1.1--10L/min
-Flow Rate of Rapid Supply: 25--75L/min
-Vaporizer: Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer
-Respiratory Circuit: Working mode: all-close, semi-close, semi-open
-Releasing pressure: 0.5-7kPa
-Respiratory Bellows: Bellows for adults, bellows for children, tidal volume range: 0-1500ml -Display Mode: High-definition 5.7” LCD screen display
-Ventilation Mode: IPPV, MANUAL (Manual monitoring of tidal, volume, ventilation volume, respiratory rate)
-Ventilation Tidal volume: 50-1500ml
-Specifically designed for surgical locations where space or access is limited
-Working mode: Manual
-Patient scope: from Pediatric to Adult (3kgs over)
-Tidal volume: 20-1500ml
-Working pressure: O2: 0.3—0.4Mpa, N2O:0.3—0.4Mpa
-Flowmeters:2 tube 2 gas flowmeter
-O2 range: 0.1-10L/min
-N2O range: 0.1-10L/min
-O2 and N2O linkage device (ORC system), FiO2 ≥25% -APL valve adjustment range: 0.5-6.0Kpa
-O2 Flush: 30-60L/min